Cosmetic Condition: The instrument is in very good cosmetic condition but displays "High Limit Remove From Service" error message when powered on. This occurs due to the following:
"SAFETY CONTROLS FOR PROTECTION FROM HIGH TEMPERATURE The microprocessor controls the temperature when the water in the BLANKETROL III equipment reaches 42.0°C ? 0.5°C (107.6°F ? 1.0°F). If water in the BLANKETROL III equipment reaches 44.0°C ? 2.0°C (111.2°F ? 3.6°F), the safety device shuts off the unit, the Status Display flashes HIGH LIMIT REMOVE FROM SERVICE and the trouble alarm sounds. If software fails and water in the BLANKETROL III equipment reaches 44.0°C ? 2.0°C (111.2°F ? 3.6°F), the back-up safety device shuts off the unit, Status Display flashes HIGH LIMIT REMOVE FROM SERVICE, and the trouble alarm sounds. If the reservoir water temperature reaches 46.0°C ? 2.0°C (114.8 ? 3.6°F), the thermostatic snap disc shuts down the unit, Status Display flashes SNAP DISC REMOVE FROM SERVICE, the patient display indicates EE02, and the trouble alarm sounds"